Project Description

Heated Sample Gas Line

GO-H 300 A

Sampling- Processing- Analysis

Heated Sample Gas Line

GO-H 300 A

Sampling- Processing- Analysis

HEATED SAMPLE GAS LINE GO-H 300 A

Sampling– Processing– Analysis

Heated measuring gas line with flexible hose line to maintain the gas temperature, as well as to prevent condensation.

Beheizte Messgasleitung GO-H 300 A Gasanalytik

THE GO-H 300 A AT A GLANCE

  • Vermeidung von Kondensatbildung Avoid condensation

  • Replaceable PTFE core

  • Moisture-protected, thermally insulated and surrounded by polyamide heat conductors

Beheizte Messgasleitung GO-H 300 A Gasanalytik

TECHNICALLY MATURE AND EXPANDABLE

Get accurate results with the GO-H 300 A

Standard equipment

  • Temperature sensor PT 100 (other possible)
  • Connection cable 1,5 m
  • Heating conductor construction in accordance with DIN
  • Production lenght 1 – 50 m

Options / accessory

  • Supplied power supply line for heated gas sampling probe
  • Sample gas hose lining DN 02
  • Sheathing mesh made of steel or VA
  • Corrugated hose made of PA or VA

THE PERFORMANCE PROFIL GO-H 300 A

Nominal temperature max. 200°C (other options on request)
Lining DN 04 or DN 06
Gas connection Lining protrudes 100 mm
Heating wire Construction according to DIN, water proof with sheathing mesh to 250 °C
Temperature sensor Pt100
Length max. 50 m
Power supply 230 V / 50 Hz
Power consumption approx. 100 Watt/Meter
Mark of conformity SEV TP 20 B/3 A. 1982
DOWNLOAD DATA SHEET
BACK TO THE OVERVIEW GAS ANALYTICS

THANK YOU FOR YOUR INTEREST IN THE GO-H 300 A

Do you have anymore questions? Do you want to get to know the GO-H 300 A?

Download the product data sheet or contact us personally. We are glad to be here for you! We will be glad to help you.

PLEASE CONTACT US