Cooling. Filter. Drying.

Gröger & Obst GOT-200

Achieve optimum preparation of the gas to be analyzed with the GOT measurement gas processing system. Effective combination of a Peltier cooler with cyclone heat exchanger, filters and permeation dryers, which reliably removes dust and dries the measurement gas.

GOT 200

Achieve optimum preparation of the gas to be analyzed with the GOT measurement gas processing system. Effective combination of a Peltier cooler with cyclone heat exchanger, filters and permeation dryers, which reliably removes dust and dries the measurement gas.

The measurement gas processing systems of the GOT 100 and GOT 200 series offer, as a complete solution, an effective combination of a Peltier cooler with cyclone heat exchanger, filters and permeation dryer that reliably remove dust and dry the sample gas.

In addition, aerosols and contaminations are retained. The resulting output dew point is below -10 °C and ensures that no further condensation is to be expected in the gas analyzer.

Facts

Benefits

  • Aerosols and contaminations are retained
  • Reliable dust removal and drying of the measurement gas
  • One unit for cooling, filtering and drying
  • Clearly arranged preparation on a mounting plate
  • Ready to operate in approx. 10 min

Equipment options

  • Special power plug for the heated gas extraction sensor
  • Special power plug fot the temperature regulator of the analysis line
  • Mounting bracket for heated analytical line for attachment to the case
  • Diaphragm filter with condensate monitor to shut off the measured gas pump if moisture ingresses
  • GO-M configuration for installation in analysis cabinets

The performance profile of the GOT-200

  • Cooler
    Peltier cooler GO-PK 1 or GO-PK 2 with cyclone heat exchanger, function-monitored Peltier element, potential-free temperature alarm contact
  • Ambient temperature
    max. 30 °C
  • Cooling temperature
    regulated at +5 to +8 °C
  • Cooling capacity
    110 kJ/h
  • Flow rate
    50 to 200 l/h
  • Input dew point
    max. 70 °C
  • Output dew point
    < -10 °C, depending on inlet dew point and flow rate
  • Condensate drain
    2 pumps, conveying performance 0.5 l/h each,
    collector 1 l
  • Filters
    aerosol filter, membrane filter 10 μ
  • Drying
    peltier element and permeation dryer
  • Gas connections
    PP bulkhead fittings 6 x 8 mm
  • Volume of the way of gas
    approx. 200 ml
  • Materials conducting sample
    glass, PP, PVC, PTFE, PVDF acrylic
  • Ready to operate
    within approx. 10 minutes
  • Power input
    max. 950 VA (depending on the length of the heated pipe)
  • Mains connection
    230 V, 50 Hz